Advances in bone healing: progress with Ti-Mg hybrid implants
Electron micrograph of the interface between magnesium (Mg) and titanium (Ti) part, including PEO coating. It can be seen how the coating extends over both materials
Image: Hereon / M. Fazel
Hybrid implants, combining a dissolvable part and a permanent part, show promise for improving bone healing after accidents or illnesses. The permanent part provides stabilization, while the dissolvable part compensates for bone defects and promotes regeneration. Titanium and magnesium are particularly suitable materials for this purpose.
Previous experiments at the Institute have demonstrated that sintering techniques effectively join these materials. However, the accelerated degradation of magnesium due to its bond with titanium remains a challenge. In collaboration with the Institutes of Surface Science and Materials Physics, a potential solution has been identified: applying a plasma electrolytic oxidation (PEO) layer to a sintered Mg-Ti specimen. This coating evenly covers and protects both parts, significantly reducing degradation without completely preventing dissolution. High-resolution synchrotron tomography revealed that the PEO layer's internal structure depends on the underlying material. Further experiments are needed, but these promising results represent a step towards hybrid implants.
Fazel et al., Journal of Magnesium and Alloys 2024 12(8):3142-3158